Features
• Looppowered4-20mAoutput
• Compactdesign
• Combineswith4-20mApHprobe
forpHcompensation
• Automatictemperature
compensation
• Optionalowcellandowmeter
insuresconstantpressureforaccurate
measurement
• Inexpensivereplaceable
membraneandelectrolyte
Description
100SchoolStreet
Andover,MA01810
978-749-9949
Tollfree-855-747-7623
www.Aquametrix.com
Applications
• Drinkingwaterdisinfection
• Wastewaterdisinfection
• Chlorinegenerator
• Swimmingpoolsandspas
Model AM-FCL Series
4-20 Direct Output Free Chlorine Sensor
by
Water Analytics
Metrix
by
Water Analytics
Metrix
TheAM-FCLisaverycompactfreechlorinesensor.Itoperateson
membrane-basedamperometrictechnology,whichprovidesthemost
trouble-freeoperationofallreagent-freemethods.Thesensoroutputsa
direct4-20mAsignalsothereisnoneedforaseparatetransmitter.The
signalistemperaturecompensated.Simplyinserttheprobe’stwowires
intoyour4-20loopandyou’regoodtogo.Combinetheoutputofthe
AM-FCLwiththatofanAquametrix4-20mAoutputprobeandyourPLC
canbeeasilyprogrammedtoperformautomaticpHcompensation.
Maintainingconstantpressureonthemembraneisessentialforaccu-
ratereading.Theoptionalowcellmaintainsthepressureandanother
optionalowmeterinsuresatightertolerance.
Thesensorisavailableinthreeranges:0to2ppm,0to5ppmand0to
10ppm.
WithAM-FCL-FC ow cell
AM-FCL-FM ow meter