Model 2200R Page 19
5.15 Status
5.15.1 The 2200R continuously checks the integrity of all stored data and monitors the
condition of the measuring system. If a fault is detected, the STATUS LED above
the display will change from green to red. When this occurs, enter the menu and
scroll through the menu until the STATUS LED illuminates. The display will show a
numerical code to assist you, in correcting the problem.
The following table shows the display codes, causes and remedies:
ORP probe efficiency <60% or
>110% Nernst coefficient (Probe
slope out of tolerance
Clean probe and recalibrate. Salt
bridge / probe may need
replacement
Temperature Sensor off Scale
Check for open or short
connections of temperature sensor
input (Terminals BK & YL on TB2)
Verify process. Check for open or
short connections
Call your AquaMetrix Rep. Or
Water Analytics directly
Call your AquaMetrix Rep. Or
Water Analytics directly
Check scale setting (See range
change in Utility Menu)
Factory settings in force as a result
of the “Factory Reset Procedure”
ORP calibration not correct, or not
performed
Re-calibrate (See Sec. 5.2 & 5.3)
Temperature calibration not correct
or not performed
Calibrate temperature (See Sec.
7.4)
NOTE: Code 4 and 5 could be serious failures so the alarm relay will activate, in addition to the
status LED changing to red, if programmed by the position of DIP switch No. 5 of Bank S1.
Refer to Section 4.7.