MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Section 2 - Installation
Section 2 - Installation Page 3
2.1 Unpacking
Save the shipping carton and packing material in case the instrument needs to be stored or
returned. Inspect the instrument and packing material for shipping damage and report any
problems immediately.
2.2 Location
Locate the tr
ansmitter close to the sensor. The list below gives typical maximum distances for
various sensors. Refer to the sensor specifications for exact information.
• Aquametrix Differential PH Probe 3000 ft (914 meters)
• Aquametrix Combination PH Probe 10 ft (3 meters)
• Aquametrix Conductivity Probe 300 ft (91 meters)
• Aquametrix Flow sensor 2000 ft (610 meters)
2.3 Mounting
Fig 2.1 SHARKTX Transmitter dimensions
Dwg# N106-128